Key Features :
Applications :
Introduction :
VEER make Digital Iron Loss Tester VDW-2065 is portable & easy to use Instrument which is very useful to find Watt loss of Electrical Steel immediately. Stamping, Laminations can not be tested on Epstein Frame because of size. So, we have designed a direct reading type Single Sheet Tester called Digital Iron Loss Tester VDW- 2065. It provides very fast testing of final products like Stamping, EI laminations etc.
Theory of Operation :
The Probe is provided with two coils. One coil is for excitation & second one is for detection of Magnetic Flux. When we put Probe on Test Specimen, it completes magnetic path. Watt meter measures iron loss of specimen & you can see Iron Loss (Watt/kg or Watt/lb)immediately on 7 segment Red Digital Display on front panel. You can set Thickness up to 0.65mm & Induction up to 1.8 Tesla (18000 Gauss) using respective setting Knob.
Specification :
Parameter | VDW-2065 (Standard) | VDW-2065E (Economy) |
Measuring Range | 0.2 to 19.00 Watt / kg or Watt / lb | |
Test specimen Thickness | 0.2 to 0.65 mm | |
Flux Density | 1.0 to 1.8 Tesla | 1.5 Tesla only |
Accuracy | ± 5% rdg ± 0.3% mg (For 10 Sample Avg Reading) @ 1.5 T for standard sample | |
Measuring Frequency | 50 / 60 Hz Internally set by default Factory. | |
Material to be Test | CRNO, CRGO, HR, CRC etc. | |
Power supply | AC 230V ± 10% @ 50 Hz / AC 110V ± 10% @ 60 Hz (on request) | |
Size of Specimen | 25mm x 70mm min. required | |
Size of Instrument (WDH) | 253mm X 300mm X 123 mm approx. | |
Weight of Instrument | 4.5 kg approx. | |
Warranty | 1 year against manufacturing defects & Extended optional. No warranty for accessories. |